ACURA MANUAL - Microdispenser pipettes |
|
 | Acura® 865 microdispensing pipettes Hand-held microdispenser intended for repeat distribution of microliter volumes. Includes a three-way valve system for automatic self-refilling. Carefully selected materials ensure durability and optimal resistance to aggressive media. Three models cover the range from 5 to 1000 µL. - Excellent ergonomics
- Ultra soft plunger activation
- Large volume display
- Versatile feeding through bottle, tubing or syringe
- Easy in-lab calibration
- Individually serial numbered and Q.C. certified
- Autoclavable fully assembled at 121°C
Three-year warranty | Ordering and performance information| | | | Inaccuracy (Rel. err.%) | Imprecision (CV %) | | Cat. No. | Vol. µL | Div. µL | Min. vol. | Mid. vol. | Max. vol. | Min. vol. | Mid. vol. | Max. vol. | | 865.0050 | 5 - 50 | 0.1 | <± 5.0% | <± 3.5% | <± 1.5% | < 2.0% | < 1.4% | < 0.4% | | 865.0200 | 20 - 200 | 0.2 | <± 2.5% | <± 1.8% | <± 1.0% | < 1.5% | < 1.0% | < 0.3% | | 865.1000 | 100 - 1000 | 1 | <± 1.5% | <± 1.1% | <± 0.6% | < 0.6% | < 0.4% | < 0.2% |
| Performance values obtained with bi-dist. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with EN ISO 8655. |
|
|